Installation
Install the LSV option according to the following procedure:
•
Replace the standard installed tubing and needles with the
tubing and needles of the LSV option. Follow the procedure
described in
Maintenance/Service
, page 29.
•
Replace the standard tray by the LSV tray.
•
Install the 500 µL loop to port 1 and 4.
•
Check the needle penetration depth and if necessary adjust
it. See page 29,
Needle Pair Replacement
.
•
Connect the drain to port 3 if the loop will be filled on a time
basis by headspace pressure. It is also possible to fill the
loop with the DDS. This will take less sample, but more time.
NOTE:
If the loop is filled using the DDS, a larger preflush volume (minimum
50 µL) must be programmed to avoid cross contamination due to the
larger internal diameters of the tubing and needle.