The first line shows that the second injection from vial 15 is
being processed, vial 82 is the last vial to be processed. The
temperature of the cooling tray shown at the end of the first line
will only be displayed if the option is present.
The second line shows the time since the injection in minutes,
seconds format, and the status of the auxiliaries. The number
will be displayed if the auxiliary is ON (only shown when enabled
in the settings).
The first character of the second line will be a blinking
R
if the
ProStar 400 is under remote control or a blinking
p
if a priority
sample is programmed.
It is possible to reprogram RUN PARAMETERS during RUN;
simply press PROG/END to enter the RUN PARAMETER mode.
Press PROG/END to exit RUN PARAMETER mode and to
continue the RUN program.
NOTE:
If you do not return to RUN mode, RUN will stop after finishing the
current injection.
Changes in the RUN parameters will be executed immediately
after returning to RUN mode, although an injection will not be
interrupted.
COOLING CONTROL can also be reprogrammed during RUN;
press F,7 to enter COOLING CONTROL mode.
Return to RUN (F,0) to continue RUN.
HOLD
HOLD (press HOLD/CONT) can be used to freeze the time
base. During HOLD, the RUN PARAMETER mode, COOLING
CONTROL mode and PRIORITY mode are accessible to
reprogram/check parameters. Exit any of these modes (if
entered) and press HOLD/CONT key to continue RUN. Analysis
time of the current injection has then been extended with the
time that HOLD was active.