990 CLD Autoline™ Leak Detector
DR
AFT 2/2/04
4-15
Troubleshooting
This section is a general guideline to help diagnose and correct common problems that may
be encountered on 990 CLD-based leak testing systems. For additional assistance on
specific 990 CLD leak detector issues, please contact Vacuum Technologies.
General
Start by recording all symptoms encountered, including any quantitative data. Be sure to
keep a calibration and maintenance log, and keep it current.
Prior to the discussion of the symptoms listed below, it is assumed that you have already
verified some very basic corrective actions that apply to all of the symptoms:
❑
The power cord is connected.
❑
All cables between the 990 Controller and the turbo pump and spectrometer tube are
properly installed, and that the cable clamps at the spectrometer tube are in place.
❑
All connections between the 990 Controller and the testing station are properly
installed.
❑
If applicable, that the oil level and color in all mechanical pumps has been checked
or that the oil was recently changed.
❑
All OPTO-22 input modules are firmly attached to the connectors at the rear of the
990 Controller.
❑
Re-check that all controls for calibration and that set points are in their proper
positions. Consult your maintenance log and setup processes for this information.