990 CLD Autoline™ Leak Detector
DR
AFT 2/2/04
4-17
Symptom Checklist
❑
“High Vacuum Pump Ready Light Does Not Illuminate”
❑
“Spectrometer Tube Pressure is High (above the green band)” on page 4-18
❑
“Filament Does Not Turn On” on page 4-18
❑
“There Is No Helium Response” on page 4-18
❑
“Station Fails to Pump Down While Displaying Gross Vacuum Failures” on
page 4-20
❑
“There is an unusually High Number of Rejects” on page 4-20
❑
“Miscellaneous Problems” on page 4-21
High Vacuum Pump Ready Light Does Not Illuminate
1. Verify that the turbo fore pump is powered, running, and producing a foreline pressure
less than 5 Torr (5000 milliTorr) while the leak detector is operating in the low sensitivity
mode (maximum is 1 Torr when in the high sensitivity mode). This applies to the
Vacuum Technologies Model V70D MacroTorr pump only.If the system contains a
Vacuum Technologies Model V70 Turbo pump, the forepressure must be less than 0.5
Torr (500 milliTorr) while the leak detector is operating in the low sensitivity mode
(maximum is 200 milliTorr when in the high sensitivity mode). If it is greater than either
value, the turbo pump or controller may trip off on an overload. To resume operation,
determine and correct the cause of high forepressure, then turn the controller off then
back on to reset the circuit.
2. With the power on:
Foreline pressure should be:
❑
≤
5 Torr for low sensitivity for the V70D
❑
≤
1 Torr for high sensitivity for the V70D
❑
≤
.5 Torr for low sensitivity for the V70
❑
≤
.2 Torr for high sensitivity for the V70
3. Check the HI VAC PUMP fuse. If blown, it may be safely replaced with a 3 A Slo-Blo
fuse for 120 V units or a 1.5 A fuse for 240 V units.
4. Check for a disconnected or damaged turbo pump cable.
NOTE
Allow up to 45 minutes for a newly-installed pump to reach full
speed (the HIGH VACUUM PUMP READY light is lit). If the
light fails to light, replace the 990 Controller to determine if the
controller is at fault. Replace the pump if a new Controller
doesn't solve the problem.