Rear panel components are shown in Figure 2-5.
v
Figure 2-5
959D Rear Panel Components
➀
Exhaust fans
Two exhaust fans cool the high vacuum pump and internal elec-
tronic assemblies.
➁
Speaker
A grid-covered speaker amplifies an audible signal present during
leak testing.
➂
Roughing/Foreline
Connection
An NW-25 connection to the TriScroll 610 pump.
➃
Power cable
120 or 220 VAC power cable, depending on the configuration. The
power configuration is factory-set.
➄
Phono Jack
(under manifold)
Connection for an optional loudspeaker or headphones and an
optional rate meter or chart recorder.
➅
Circuit Breakers
The POWER circuit breaker provides AC to the leak detector and to
the TriScroll 610 (see Figure 1-2 on page 1-6). The MAIN POWER
switch inside the front panel access door subsequently turns the
leak detector on or off.
The ELECTRONICS circuit breaker provides independent protec-
tion to the power distribution within the leak detector.
Both breakers are push-to-reset.