959D Leak Detector
1-10
DRAFT
9/13/02
1.6 Internal Valving
The vacuum system is shown in Figure 1-4.
Figure 1-4
Vacuum System
1.6.1 Direct Operation of the Valves
The 959D valve system allows for evacuation, testing, and venting the object to be tested.
The main electronics are responsible for automatically sequencing the testing operation as
described in the next section. The front panel operating buttons display the current leak
detector mode as well as function as operation buttons. All valve coils operate at line volt-
age. There are four valves:
❑
Roughing valve (V1)
❑
Test valve (V2)
❑
Vent valve (V3)
❑
Isolation valve (V6)
The V6 isolation valve is normally closed when power is OFF, and opens
automatically upon the application of electrical power.