73
7532
Maintenance
7
Maintenance
The 7500 RTG measuring instrument requires no special maintenance.
Exterior cleaning
When cleaning the exterior of measuring devices, always use cleaning agents that do not
attack the surface of the housing and the seals.
Replacement
After a complete 7500 RTG or electronic module has been replaced, the parameters can
be downloaded into the instrument again via the communication interface. Prerequisite
to this is that the data were uploaded to the PC beforehand using the ToF Tool /
Commuwin II.
Measurement can continue without having to carry out a new setup.
• You may have to activate linearization
• You may need to record the tank map again
After an antenna component has been replaced, a new calibration must be carried out.