15
7532
Mounting
Measuring range
The usable measuring range depends on the size of the antenna, the reflectivity of the
medium, the mounting location, and eventual interference reflections.
The following tables describe the groups of media as well as the achievable measuring
range as a function of application and media group. If the dielectric constant of a
medium is unknown, it is recommended to assume media group B to ensure a reliable
measurement.
1)Treat Ammonia NH3 as a medium of group A, e.g. always use a stilling well.
Measuring range depending on product class for 7532 RTG:
Product class
DK (
ε
r)
Examples
A
1.4 … 1.9
non-conducting liquids, e.g. liquefied gas
1)
B
1.9 … 4
non-conducting liquids, e.g. benzene, oil, toluene, …
C
4 … 10
e.g. concentrated acids, organic solvents, esters, aniline,
alcohol, acetone, …
D
> 10
conducting liquids, e.g. aqueous solutions, dilute acids and
alkalis
Product class
Stilling well/ Bypass
Measuring range
7532 RTG
> DN150
A
DK(
ε
r)=1,4…1,9
38 m/124 ft
B
DK(
ε
r)=1,9…4
38 m/124 ft
C
DK(
ε
r)=4…10
38 m/124 ft
D
DK(
ε
r)>10
38 m/124 ft
max. measuring
range with custody
transfer approvals
20 m/65 ft
Summary of Contents for 7532
Page 31: ...31 7532 Operation 5 Operation 5 1 Quick operation guide...
Page 44: ...Operation Radar Tank Gauge 44 Installation and Operations Manual...
Page 46: ...Commissioning Radar Tank Gauge 46 Installation and Operations Manual 6 3 Basic Setup...
Page 72: ...Commissioning Radar Tank Gauge 72 Installation and Operations Manual...
Page 74: ...Maintenance Radar Tank Gauge 74 Installation and Operations Manual...
Page 76: ...Accessories Radar Tank Gauge 76 Installation and Operations Manual...
Page 77: ...77 7532 Troubleshooting 9 Troubleshooting 9 1 Troubleshooting instructions...
Page 91: ...91 7532 Appendix 11 Appendix 11 1 Operating menu HART Display modul ToF Tool...
Page 105: ...7532 IOM027FVAE0804 105 Notes...