Standards and IEC Classifications
4-10
LigaSure Vessel Sealing Generator Service Manual
Voltage Transients (Emergency Generator Mains
Transfer)
This generator operates in a safe manner when you transfer between line AC and
an emergency generator voltage source.
Guidance and manufacturer's declaration - electromagnetic emissions
The LigaSure is intended for use in the electromagnetic environment specified below. The customer or the user of the
LigaSure should ensure that it is used in such an environment.
Emissions test
Compliance
Electromagnetic environment -
guidance
RF emissions
CISPR 11
Group 1
The LigaSure uses RF energy only for
its internal function. Therefore, its RF
emissions are very low and are not
likely to cause any interference in
nearby electronic equipment.
RF emissions
CISPR 11
Class A
The LigaSure is suitable for use in all
establishments other than domestic
and those directly connected to the
public low-voltage power supply
network that supplies buildings used
for domestic purposes.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage fluctuations/ flicker
emissions IEC61000-3-3
Complies
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com