High Voltage (HV) Power Supply Board
8-6
LigaSure Vessel Sealing Generator Service Manual
4.
Remove four cables from the HV power supply board:
*All reference points in this table appear on the HV power supply board.
Two latches secure the ribbon cable connector to the HV power supply board.
Release them by rotating them away from the connector, then pull the
connector away from the board.
5.
Remove three screws (M4 x.7 x 8) that secure the HV power supply board to
the chassis.
Note that the screw nearest the rear of the unit also fastens the chassis ground
to the board by way of a ring tongue terminal.
6.
Lift the HV power supply board out of the chassis.
Figure 8-3.
Heat sink bracket and cable
connections on the HV power
supply board
Name
Ref*
Location
Wires
Colors
Switched AC
power cable
J2
upper rear corner
of the board
3
brown, blue, and
green/yellow
AC switch #1
cable
J1
next to the first
cable
2
brown and blue
HV power
supply control
ribbon cable
J3
near the front
edge of the board
ribbon
red and gray
HV DC cable
J4
under the HV
power supply
control ribbon
cable
4
two black and two
red
Switched AC
Power Cable
AC Switch
Cable
Heat Sink Bracket
HV Power Supply Control
Ribbon Cable
HV DC Cable
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com