Replacing the Battery
8-12
LigaSure Vessel Sealing Generator Service Manual
Install the Control Board
Whether you have replaced the battery or the entire board, follow these steps to
install the control board:
1.
Verify that the packaging for the new control board contains the part number
for the LigaSure control board. Do not install any other control board.
2.
Install the battery in the battery socket (positive side facing up).
3.
Install the control board:
a.
Position the control board over the RF board with the components facing
the left panel.
b.
Slide the board down, carefully fitting the edge connector into the
matching connector on the RF board.
4.
Position the mounting brackets, one on each upper corner and one on the
upper center of the board. Replace three screws (M4 x .7 x 8) to secure the
mounting brackets and the control board to the chassis.
5.
Connect the display board ribbon cable to the control board and lock the
connector.
6.
Connect the power cord to the wall receptacle. Turn on the generator.
7.
Calibrate the generator. Refer to Chapter 6,
Calibrating the LigaSure
Generator
for further instructions. If the calibration is successful, go to the
next step.
8.
Turn off the generator. Disconnect the power cord from the wall receptacle.
9.
Install the cover. For further information, refer to
Cover
in this chapter.
10.
Connect the power cord to the wall receptacle.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com