Performance Characteristics
4-6
LigaSure Vessel Sealing Generator Service Manual
Low Frequency (50-60 Hz) Leakage Current
(AAMI HF-18-1993)
High Frequency (RF) Leakage Current (IEC 60601-2-2)
Enclosure source
current, ground open
< 300 µA
Source current, patient leads, all outputs:
Normal polarity, intact
ground
< 10 µA
Normal polarity,
ground open
< 50 µA
Reverse polarity,
ground open
< 50 µA
Sink current at high
line, all inputs
< 10 µA @ 120 V
< 50 µA @ 264 V
Bipolar RF leakage
current
≤
69 mA rms
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com