Inspecting the Generator
T
roubleshooting
LigaSure Vessel Sealing Generator Service Manual
7
-
3
4.
Check the bipolar handset receptacle on the front panel for obstructions or
damage. You can connect either a footswitching or handswitching bipolar/
macrobipolar handset to this receptacle. Insert the handset plug firmly into
this receptacle to verify a secure fit.
If the connection is loose, try another handset. If the connection is still loose,
replace the front panel. Refer to Chapter 8,
Front Panel and Display Board
for
further instructions.
5.
Check the vessel sealing handset receptacle on the front panel for obstructions
or damage. Connect a footswitching vessel sealing handset to this receptacle,
inserting the plug firmly into the receptacle to verify a secure fit.
If the connection is loose, try another handset. If the connection is still loose,
replace the front panel. Refer to Chapter 8,
Front Panel and Display Board
for
further instructions.
6.
With the bipolar and vessel sealing handsets properly mated, check that the
smart indicator lamp illuminates green. If the smart indicator lamp illuminates
red, use the following procedure:
a.
Verify that you have connected the correct LigaSure handset to the seal
receptacle.
b.
Verify that the LigaSure handset fits securely in the seal receptacle.
c.
Verify that you have connected a smart bipolar adapter to the bipolar
receptacle.
d.
Verify that the smart bipolar adapter fits securely in the bipolar receptacle.
7.
If the smart indicator lamp remains red, replace the front panel. Refer to
Chapter 8,
Front Panel and Display Board
for further instructions.