Front Panel and Display Board
Replacement Procedures
LigaSure Vessel Sealing Generator Service Manual
8
-
19
Figure 8-12.
The display board positioned on
the front panel
Remove the Display Board
The display board resides in the back of the front panel. Removing the board
involves disconnecting three cables, removing five screws, and lifting the board
out of the front panel.
1.
Disconnect three cables:
Latches hold the smart input/output (I/O) cables in their connectors. To release
the cable, rotate the latches away from the connector.
2.
Remove five screws (4-20 x 5/16 in. thread forming):
•
Two screws on either end of the board.
•
One screw immediately under the ribbon cable.
When you remove the two upper corner screws, you also remove left and
right display board ground clips. Be sure to put these screws and clips
together in a safe place.
3.
Lift the display board out of the front panel.
Name
Ref
Location
Wires
Colors
Smart I/O
cables (2)
RF board
red, gray
Flex ribbon
cable
J3 on
display
board
Near right edge of
display board
translucent, with
metal traces
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com