Responding to System Alarms
7-14
LigaSure Vessel Sealing Generator Service Manual
108
CORRECTED_RTC_CRC_DATA
Real time clock RAM CRC test failed
during initialization, but the data was
restored from battery backed RAM.
Reported to serial port only.
Replace the battery.
If the error code reappears, replace the control board (failed
U1).
109
CORRECTED_BATRAM_CRC_DATA
Real time clock RAM CRC test failed
during initialization. An attempt to correct
the problem failed.
Replace the battery.
If the error code reappears, replace the control board.
110
111
112
113
114
INVALID_RECENT_ERRORS
Recent error data kept in battery backed
RAM was found to have an invalid CRC
at powerup.
INVALID_ERROR_COUNTS
Error counts kept in battery backed RAM
was found to have an invalid CRC at
powerup.
INVALID_LAST_POWER_SETTINGS
Last power settings structure in battery
backed RAM was found to have an
invalid CRC at powerup.
INVALID_STATS_COUNTS
Statistical data kept in battery backed
RAM was found to have an invalid CRC
at powerup
INVALID_MINUTES_ON_WARNING
Minutes On data kept in battery backed
RAM was found to have an invalid CRC
at powerup.
Reported to serial port only.
Replace the battery.
If the error code reappears, replace the control board.
115
MM_DOSERR_TEST_ERROR
Master microprocessor reported a
powerup DOSERR test failure.
Turn the power switch off (O) then on again (|).
If the error reappears, replace the control board.
If the error reappears, replace high voltage power supply.
If the error reappears, replace RF output FET Q5 on the RF
board.
116
DOSERR_TEST_MEMORY_FAULT
Doserror test state variable corrupted
during powerup dose error test.
Turn the power switch off (O) then on again (|).
If the error reappears, replace the control board.
117
MICRO_COMPATIBILITY_ERROR
Incompatible software versions on
feedback and master processors, or
master side software incompatible with
digital board type.
Turn the power switch off (O) then on again (|).
If the error reappears, replace the control board.
118
DPRAM_RENDEZVOUS_ERROR
Master and feedback processors fail to
communicate on power up. Either timing
or content of communication invalid.
Turn the power switch off (O) then on again (|).
If the error reappears, replace the control board.
Number
Description
Recommended Action