Footswitch Receptacles
Controls, Indicators,
and Receptacles
Force FX-8C Service Manual
2-9
Footswitch Receptacles
The rear panel contains three footswitch receptacles: two for monopolar
and one for bipolar.
Monopolar Footswitch Receptacles
You must connect a monopolar footswitch if you connect a monopolar
footswitching instrument to the generator.
Connect a two-pedal monopolar footswitch to the Monopolar 1
Footswitch receptacle.
The connected footswitch activates monopolar output for the instrument
that is connected to the Monopolar 1/CEM Instrument receptacle on the
front panel.
Connect a two-pedal monopolar footswitch to the Monopolar 2
Footswitch receptacle.
The connected footswitch activates monopolar output for the instrument
that is connected to the Monopolar 2 Instrument receptacle on the front
panel
Bipolar Footswitch Receptacle
You must connect a bipolar footswitch if you connect a bipolar
footswitching instrument to the generator.
Connect a single-pedal bipolar footswitch to the Bipolar Footswitch
receptacle.
The connected footswitch activates bipolar output for the instrument that
is connected to the Bipolar Instrument receptacle on the front panel.