Correcting Malfunctions
Troubleshooting
Force FX-8C Service Manual
6-5
Generator is on, but did
not complete the self-test
An alarm condition exists
Check the display for an alarm number. Note the number
and refer to Responding to System Alarms in this
section.
Software malfunction
Turn off, then turn on the generator.
Loose or disconnected internal cables
Check and correct all internal connections.
Faulty low voltage power supply
Check the low voltage power supply.
Damaged Control board connectors and/
or malfunctioning Control board
Remove the Control board and inspect the connectors to
the Power Supply/RF board and to the Display board for
damage, poor seating, etc. If the problem persists,
replace the Control board. For instructions, refer to
Control Board Replacement in Section 7.
Shorts or disconnects on Power Supply/
RF board
Check the Power Supply/RF board for shorts or
disconnects.
Faulty power switch
Replace the power switch. Refer to Panel Power Switch
Replacement in Section 7.
Malfunctioning front panel components
Replace the front panel assembly. Refer to Front Panel
Replacement in Section 7.
Alarm number 212
appears in the Cut
display during the self-
test
The battery was removed and/or replaced,
but the generator was not calibrated
Turn off, then turn on the generator to clear the number.
Calibrate the generator. For instructions, refer to
Calibrating the Generator in Section 5.
Faulty battery on Control board causing
loss of battery-backed memory
Replace the battery and check the battery-backed RAM
device. Refer to Correcting Battery-Backed RAM
Malfunctions later in this section.
Loss of battery-backed RAM due to faulty
component on the Control board (other
than the battery)
Replace the Control board. Refer to Control Board
Replacement in Section 7.
Activation and/or alarm
tones do not sound;
speaker is
malfunctioning
Poor connection or damaged footswitch
board ribbon cable
Check/correct connection. If indicated, replace the
Footswitch board. Refer to Footswitch Board
Replacement in Section 7.
Faulty connections or speaker on
Footswitch board
Replace the Footswitch board. Refer to Footswitch Board
Replacement in Section 7.
Audio signal malfunction on Control board
Replace the Control board. Refer to Control Board
Replacement in Section 7.
Situation
Possible Cause
Recommended Action
Summary of Contents for Force FX-8C
Page 1: ...Service Manual Force FX 8C Electrosurgical Generator with Instant Response Technology...
Page 30: ...Notes 2 12 Force FX 8C Service Manual...
Page 56: ...Notes 3 26 Force FX 8C Service Manual...
Page 58: ...Block Diagram 4 2 Force FX 8C Service Manual Block Diagram...
Page 84: ...Notes 4 28 Force FX 8C Service Manual...
Page 156: ...Interconnect Diagram 7 2 Force FX 8C Service Manual Interconnect Diagram...