Number
Description
Recommended Action
222
223
Main microcontroller failed to copy
battery-backed RAM data to dual-port
RAM in time allotted
Main microcontroller failed to complete
initialization in time allotted
Refer to Correcting IC U3 Malfunctions in this section.
224
Internal diagnostics. Calibration data
checksum error on feedback
microcontroller (data in error shown on
terminal using serial port)
Refer to Correcting IC U6 Malfunctions in this section.
225
Internal diagnostics. Feedback
microcontroller CPU test failed
1. Replace the Control board. Refer to Control Board
Replacement in Section 7.
2. Calibrate the generator. Refer to Section 5 for instructions.
If the alarm number reappears, record the number and call the
Valleylab Service Center.
226
Feedback microcontroller unable to access
FEEDBACK_SEM semaphore
Refer to Correcting IC U6 Malfunctions in this section.
230
231
Software malfunction. keyed_state value
outside state range
Software malfunction. key_req_mode
value outside state range
Turn off, then turn on the generator. If the alarm number
reappears, record the number and call the Valleylab Service
Center.
232
Feedback microcontroller unable to access
KEY_ACTIVE_SEM semaphore
Refer to Correcting IC U6 Malfunctions in this section.
240
241
242
244
245
Software malfunction. activation_seq
value outside state range
Software malfunction. feedback value in
do_feedback( ) outside state range
Software malfunction. feedback value in
do_cal_feedback( ) outside state range
Software malfunction. disable_seq value
outside state range
Software malfunction. spark_state value
outside state range
Turn off, then turn on the generator. If the alarm number
reappears, record the number and call the Valleylab Service
Center.
246
247
Feedback microcontroller unable to access
FEEDBACK_SEM semaphore
Feedback microcontroller unable to access
ECON_SEM semaphore
Refer to Correcting IC U6 Malfunctions in this section.
Summary of Contents for Force FX-8C
Page 1: ...Service Manual Force FX 8C Electrosurgical Generator with Instant Response Technology...
Page 30: ...Notes 2 12 Force FX 8C Service Manual...
Page 56: ...Notes 3 26 Force FX 8C Service Manual...
Page 58: ...Block Diagram 4 2 Force FX 8C Service Manual Block Diagram...
Page 84: ...Notes 4 28 Force FX 8C Service Manual...
Page 156: ...Interconnect Diagram 7 2 Force FX 8C Service Manual Interconnect Diagram...