Correcting Malfunctions
Troubleshooting
Force FX-8C Service Manual
6-7
Generator is on and
accessory is activated,
but generator does not
deliver output
Malfunctioning footswitch or
handswitching instrument
Turn off the generator. Check and correct all accessory
connections.
Turn on the generator. Replace the accessory if it
continues to malfunction.
Footswitch connected to Monopolar 1
Footswitch receptacle is being used for
instrument connected to Monopolar 2
Instrument receptacle
Connect the footswitch to the Monopolar 2 Footswitch
receptacle.
or
Connect the instrument to the Monopolar 1/CEM
Instrument receptacle.
Footswitch connected to Monopolar 2
Footswitch receptacle is being used for
instrument connected to Monopolar 1/
CEM Instrument receptacle
Connect the footswitch to the Monopolar 1 Footswitch
receptacle.
or
Connect the instrument to the Monopolar 2 Instrument
receptacle.
Power set too low
Increase the power setting.
An alarm condition exists
Check the Cut display for an alarm number. Note the
number and refer to Responding to System Alarms later
in this section.
Blown fuse on Power Supply/RF board
Check the high voltage power supply fuse (F1) and
replace if necessary. Refer to Fuse Replacement in
Section 7.
Control board malfunction
If the indicator bar does not illuminate and the tone does
not sound, replace the Control board. Refer to Control
Board Replacement in Section 7.
High voltage power supply malfunction
(high voltage is not present during
activation)
If high voltage is not present at TP3 on the Power
Supply/RF board, troubleshoot the high voltage power
supply
RF output stage malfunction (high voltage
is present during activation)
Troubleshoot the RF output stage as described below:
On the Power Supply/RF board, verify T_ON\ pulses at
U4 during activation.
• If pulses are not present before U4, replace the
Control board. Refer to Control Board Replacement in
Section 7.
• If pulses are present at U4 but not after U4, replace
U4.
Check the power MOSFET at J9 for failure (typically fail
shorted).
Check all output relays to verify that they are toggling
during operation. If they are not, check the relay drivers
(U15 and U16).
Check for shorting of the output tuning inductors (T14
and T15).
Situation
Possible Cause
Recommended Action
Summary of Contents for Force FX-8C
Page 1: ...Service Manual Force FX 8C Electrosurgical Generator with Instant Response Technology...
Page 30: ...Notes 2 12 Force FX 8C Service Manual...
Page 56: ...Notes 3 26 Force FX 8C Service Manual...
Page 58: ...Block Diagram 4 2 Force FX 8C Service Manual Block Diagram...
Page 84: ...Notes 4 28 Force FX 8C Service Manual...
Page 156: ...Interconnect Diagram 7 2 Force FX 8C Service Manual Interconnect Diagram...