Using the RS-232 Serial Port
5-44
Force FX-8C Service Manual
Commands for Changing the Default Coag Mode
No result is displayed unless a character other than D, F, or S is entered
Then the following is displayed:
No Change
Try Again!
To save the new default mode, turn off, then turn on the generator.
Step 3 – Disconnect the Computer from the Generator
A.
Turn off the generator.
B.
Disconnect the DB 9 cable from the generator.
C.
Position the panel over the serial port and install the screws that secure
it to the rear panel of the generator.
:MC <letter>
where <letter> may be any of the following:
D
F
S
sets default coag mode to desiccate
sets default coag mode to fulgurate
sets default coag mode to spray
:MDD <value>
Modifies the selected desiccate mode setting, where <value>
represents a selection as follows:
1
2
3
Desiccate 1
Desiccate 2
Desiccate 3
:MF <value>
Modifies the selected Fulgurate mode setting, where <value>
represents a selection as follows:
0
1
LCF Fulgurate
Fulgurate