S E C T I O N
6
Force FX-8C Service Manual
6-1
6
Troubleshooting
If the generator is not functioning properly, use the information in this
section to perform the following tasks:
• Identify and correct the malfunction.
• If a system alarm number was displayed, take the appropriate action to
correct the alarm condition.
Inspecting the Generator
If the Force FX-8C generator malfunctions, check for obvious conditions
that may have caused the problem:
• Check the generator for visible signs of physical damage.
• Verify that all accessory cords are properly connected.
• Check the power cord. Replace the power cord if you find exposed
wires, cracks, frayed insulation, or a damaged connector.
• Open the fuse drawer and inspect the fuse housing and fuses for
damage and corrosion. Verify that the fuses are firmly seated
An internal component malfunction in the generator can damage the
fuses. You may need to replace the fuses if the generator fails the self-test
or stops functioning. Refer to Fuse Replacement in Section 7.