Output Characteristics
Technical Specifications
Force FX-8C Service Manual
3-13
Bipolar
Monopolar Cut
Monopolar Coag
Precise
470 kHz sinusoid
Standard
470 kHz sinusoid
Macro
470 kHz sinusoid
Low
390 kHz sinusoid. Similar to the Pure cut mode except the
maximum voltage is limited to a lower value.
Pure
390 kHz sinusoid
Blend
390 kHz bursts of sinusoid, recurring at 27 kHz intervals.
50% duty cycle.
Desiccate 1
240 kHz sinusoid repeated at 39 kHz. 8% duty cycle
Desiccate 2
393 kHz sinusoid
Desiccate 3
393 kHz sinusoid
Fulgurate
470 kHz damped sinusoidal bursts with a repetition
frequency of 30 kHz into 500 ohms
LCF Fulgurate
470kHz damped sinusoidal bursts with a repetition
frequency of 57 kHz into 500 ohms
Spray
470 kHz damped sinusoidal bursts with a randomized
repetition centered at 28 kHz. Frequencies include
21 kHz < f < 35 kHz. Output is further modulated by a
random 250 Hz envelope with a variable duty cycle.
Summary of Contents for Force FX-8C
Page 1: ...Service Manual Force FX 8C Electrosurgical Generator with Instant Response Technology...
Page 30: ...Notes 2 12 Force FX 8C Service Manual...
Page 56: ...Notes 3 26 Force FX 8C Service Manual...
Page 58: ...Block Diagram 4 2 Force FX 8C Service Manual Block Diagram...
Page 84: ...Notes 4 28 Force FX 8C Service Manual...
Page 156: ...Interconnect Diagram 7 2 Force FX 8C Service Manual Interconnect Diagram...