If the sensor surface is at too high a level, to lower the sensor, grind the surface. For
instructions on how to check the sensor level, see
Checking DRS511 Position (page 112)
1. Check that the surface temperature sensor stays above the target level.
1
Surface temperature sensor
2
Optical coverage sensor
3
Conductivity and polarizability sensor
4
Black ice sensor
5
Subsurface temperature sensor
2. If the surface temperature sensor is too high, use a screwdriver to press the metal part
of the sensor downward.
Hammer the screwdriver handle gently until the sensor lies 1 … 3 mm (0.04 … 0.12 in)
below the sensor surface.
AWS310-SITE Configuration and Maintenance Manual
M211828EN-C
114
Summary of Contents for AWS310-SITE
Page 10: ...AWS310 SITE Configuration and Maintenance Manual M211828EN C 8 ...
Page 21: ...Chapter 2 Product Overview 19 ...
Page 22: ...AWS310 SITE Configuration and Maintenance Manual M211828EN C 20 ...
Page 62: ...AWS310 SITE Configuration and Maintenance Manual M211828EN C 60 ...
Page 86: ...AWS310 SITE Configuration and Maintenance Manual M211828EN C 84 ...
Page 140: ...AWS310 SITE Configuration and Maintenance Manual M211828EN C 138 ...
Page 148: ...AWS310 SITE Configuration and Maintenance Manual M211828EN C 146 ...
Page 150: ...AWS310 SITE Configuration and Maintenance Manual M211828EN C 148 ...
Page 152: ...AWS310 SITE Configuration and Maintenance Manual M211828EN C 150 ...
Page 166: ...AWS310 SITE Configuration and Maintenance Manual M211828EN C 164 ...
Page 172: ...AWS310 SITE Configuration and Maintenance Manual M211828EN C 170 ...
Page 174: ...AWS310 SITE Configuration and Maintenance Manual M211828EN C 172 ...
Page 175: ......
Page 176: ...www vaisala com ...