5. In
Serial Port
, set the following:
COM port
The COM port to which the service cable is connected to.
In the
Default
station, the COM port is set to
1
. To make sure this matches the port
assigned for the USB serial port, launch the Vaisala USB Instrument Finder
program from the
Start
menu.
Baud rate
Set to
9600
.
Parity
Set to
None
.
Data bits
Set to
8
.
Stop bits
Set to
1
.
Flow control
Set to
None
.
6. The
Telephone
section is not needed. Leave it empty.
7. To save the changes, select
Save
.
4.3.2 Establishing Connection to Station
1. In AWS Client, select
File > Connect
.
2. To establish a connection, select the correct weather station and
OK
.
When you are connected to the station but the service connection is closed, messages and
reports appear in the main window.
The report type and appearance depend on your setup. The messages and reports only
appear if they are configured in the station.
Some values are shown as slashes, either because the sensor is missing or because the
value is calculated from the measured values later.
4.3.3 Changing COM Port
1. In AWS Client, select
Settings > Address Book
.
2. Select the appropriate station and
Edit
.
Chapter 4 – AWS Client Usage
31
Summary of Contents for AWS310-SITE
Page 10: ...AWS310 SITE Configuration and Maintenance Manual M211828EN C 8 ...
Page 21: ...Chapter 2 Product Overview 19 ...
Page 22: ...AWS310 SITE Configuration and Maintenance Manual M211828EN C 20 ...
Page 62: ...AWS310 SITE Configuration and Maintenance Manual M211828EN C 60 ...
Page 86: ...AWS310 SITE Configuration and Maintenance Manual M211828EN C 84 ...
Page 140: ...AWS310 SITE Configuration and Maintenance Manual M211828EN C 138 ...
Page 148: ...AWS310 SITE Configuration and Maintenance Manual M211828EN C 146 ...
Page 150: ...AWS310 SITE Configuration and Maintenance Manual M211828EN C 148 ...
Page 152: ...AWS310 SITE Configuration and Maintenance Manual M211828EN C 150 ...
Page 166: ...AWS310 SITE Configuration and Maintenance Manual M211828EN C 164 ...
Page 172: ...AWS310 SITE Configuration and Maintenance Manual M211828EN C 170 ...
Page 174: ...AWS310 SITE Configuration and Maintenance Manual M211828EN C 172 ...
Page 175: ......
Page 176: ...www vaisala com ...