The faults, indicated in the table below, are listed in order of priority. The faults on top of the list have
the highest priority. This means that faults with a lower priority can be overwritten by faults of a higher
priority.
At power-on the system is explicitly tested for failures indicated by display messages 1, 2, 3, 4, 5, 6
and 7. During normal operation these tests (except 2) run at the background.
下表中指出的故障系按优先顺序排列。表中上方的故障有最高优先权。这意味着拥有高优先的故障可
以覆盖低优先问题。当接通电源后,系统将明确地检测是否存在故障,此时显示器依次显示信息
1, 2,
3, 4, 5, 6
和
7
。在正常操作情况下,这些测试(
2
除外)将在本底进行。
or
or
Followed by
接下来
:
Display
message
显示信息
Viscosity
output [mA]
黏度输出
Temperature
output [mA]
温度输出
Recovery action
(attempt)
恢复动作
(
尝试
)
Fault description
故障描述
0
0
Delay & retry
15 V power failure. Supply drops below 13 V
15
伏电源故障。供电电源在
13
伏以下
0
Actual
Delay & retry
Pendulum not oscillating
钟摆无
振动
4
Actual
Delay & retry
Oscillation error
振动误差
4
4
20
4
None
None
Temperature out of range (overflow)
温度超出量程(溢出)
Temperature out of range (underflow)
温度超出量程(下溢)
20
4
Actual
Actual
None
None
Viscosity out of range (overflow)
黏度超出量程(溢出)
Viscosity out of range (underflow)
年度超出量程(下溢)
Actual
Open loop
Open loop
Actual
None
None
Temperature output open loop
温度输出开环回路
Viscosity output open loop
黏度输出开环回路
Not all segments
working after
power on
Actual
Actual
Not applicable
7 - segment display defective 7
段显示欠缺
Any other figure or letter indicates an error in the electronics or software.
所有其他数字或字母代表电子或软件误差。
Summary of Contents for VISCOSENSE 2
Page 24: ...23 Figure 12 External connections at interface box 12...
Page 37: ...36 16 CERTIFICATES Certificates are delivered separately...
Page 39: ...38 Figure 19 Dimensions of ViscoSense 2 sensor All dimensions in millimetres 19...
Page 40: ...39 Figure 20 Dimensions of ViscoSense 2 sensor All dimensions in millimetres 20...
Page 41: ...40 Figure 21 Dimensions of ViscoSense 2 sensor All dimensions in millimetres 21...