Initial use (operation)
5.3 Shutdown (switch off)
Take pumping unit out of operation
1.
Stop the process and allow the pumping unit to run for
approx. 30 minutes, with open gas ballast or open inlet (IN).
5
Condensate and media residues will be flushed out of the
vacuum pump.
IMPORTANT!
>
Prevent deposits and flush condensate out of
the pump.
2
(a)
2.
Switch rocker switch
(a)
off – switch position
0
.
5
Pumping unit switched off.
3.
Unplug the power plug.
4.
Disconnect the pumping unit from the apparatus.
5.
Empty the glass flasks.
6.
Check the pumping unit for possible damage and pollution.
Switch‑off pumping
unit