59
20901101_EN_PC5xx-6xx_select_Serie_V1.3_260819
Cleaning and maintenance
7 Cleaning and maintenance
WARNING
Danger due to electrical voltage.
>
>
Switch the device off before cleaning or maintenance
work.
>
>
Unplug the power plug from the socket.
Danger due to outdated motor condensers.
Condensers can become hot and possibly melt. There
can also rarely be darting flames.
>
>
Regularly check motor condensers (measure capac
-
ity, estimate operating hours). Typical service life is
10,000 to 40,000 operating hours, depending on op
-
erating conditions (temperature, engine load, ...).
>
>
Condensers must be replaced by an electrician.
Risk from contaminated parts.
Pumping hazardous media can result in hazardous
substances adhering to internal parts of the pump.
>
>
Wear your personal protective equipment, e.g., pro
-
tective gloves, eye protection and, if necessary, re
-
spiratory protection.
>
>
Decontaminate the vacuum pump before opening it.
If necessary have decontamination carried out by an
external service provider.
>
>
Take safety precautions according to your instruc
-
tions for handling hazardous substances.
NOTICE
Damage possible if work is performed incorrectly.
>
>
Have maintenance work performed by a trained professional
or at least by a trained person.
>
>
Recommendation: Before carrying out maintenance for the
first time, please read through all the instructions to get an
overview of the required service work.
Summary of Contents for PC 510 select
Page 12: ...12 20901101_EN_PC5xx 6xx_select_Serie_V1 3_260819 Introduction ...
Page 44: ...44 20901101_EN_PC5xx 6xx_select_Serie_V1 3_260819 Installation and connection ...
Page 90: ...90 20901101_EN_PC5xx 6xx_select_Serie_V1 3_260819 ...
Page 91: ...91 20901101_EN_PC5xx 6xx_select_Serie_V1 3_260819 ...