58
20901101_EN_PC5xx-6xx_select_Serie_V1.3_260819
Troubleshooting
Error
`
Possible cause
Remedy
Personnel
`
`
Ball bearing defec
-
tive.
`
`
Outlet pipe open.
`
`
Glass flask on EK
missing.
`
`
Diaphragm tear or
diaphragm clamping
disc loose.
`
Service the vacuum
pump and replace
defective parts or
send in the device.
`
Check exhaust gas
line connections.
`
Connect the outlet
pipe to an exhaus-
tion system, e.g.,
fume hood.
`
Assemble glass
flask.
Specialist
Condenser
(cooler) de
-
fective
`
`
Mechanically
damaged.
`
Send in.
resp. Spe-
cialist