57
20901101_EN_PC5xx-6xx_select_Serie_V1.3_260819
Troubleshooting
Error
`
Possible cause
Remedy
Personnel
`
`
Vacuum line too long
or cross-section too
small.
`
Use a shorter vac-
uum line with a
larger cross‑section.
resp. Spe-
cialist
`
`
Condensate inside
the vacuum pump.
`
Allow vacuum pump
to run for a few min-
utes with the suction
nozzle open.
Operator
`
`
Deposits inside the
vacuum pump.
`
Clean and check
pump heads.
Specialist
`
`
Diaphragms or valves
defective.
`
Replace dia-
phragms and
valves.
Specialist
`
`
High level of vapor
generated in the pro
-
cess.
`
Check process
parameter.
Specialist
`
`
High level of vapor
generated in the pro
-
cess.
`
Check settings at
the controller.
Operator
`
`
Gas ballast open
`
Close gas ballast
Operator
`
`
Gas ballast cap
porous or no longer
present.
`
Check gas ballast
cap.
`
Replace defective
components.
Operator
`
`
Pump too hot.
`
Allow the pump to
cool down. Deter-
mine and eliminate
cause of the over-
heating.
resp. spe-
cialist
No display
`
`
Pumping unit
switched off.
`
`
Power plug not cor
-
rectly plugged in or
pulled out.
`
`
VACUU·BUS plug‑in
connection or cables
defective or not con-
nected.
`
`
Controller switched
off or defective.
`
Switch pumping
unit on using rocker
switch.
`
Check power supply
and cable.
`
Check VACUU·BUS
plug‑in connection
and cables to the
controller.
`
Switch on controller.
`
Replace defective
components.
Operator
Loud operat
-
ing noises
`
`
No hose installed.
`
Check hose and
install it right.
Operator
Summary of Contents for PC 510 select
Page 12: ...12 20901101_EN_PC5xx 6xx_select_Serie_V1 3_260819 Introduction ...
Page 44: ...44 20901101_EN_PC5xx 6xx_select_Serie_V1 3_260819 Installation and connection ...
Page 90: ...90 20901101_EN_PC5xx 6xx_select_Serie_V1 3_260819 ...
Page 91: ...91 20901101_EN_PC5xx 6xx_select_Serie_V1 3_260819 ...