CO
2
MPACT ENDOSCOPIC INSUFFLATOR
®
SYSTEM Operator’s Manual
Page 34
Condition
Possible Causes
Field Checkout
“CONTINUED”
My CO
2
cylinder is
prematurely emptying.
What could be the problem?
High Pressure Hose from
CO
2
cylinder is not installed
properly.
Inspect the CO
2
cylinder Yoke and verify that the plastic disc
that creates the gas seal at the CO
2
cylinder valve is in-place
and free from defect.
After verifying all High Pressure Hose sealing surfaces are
clean and free of defect; reconnect the High Pressure Hose to
the CO
2
supply tank Yoke and CO
2
MPACT ENDOSCOPIC
INSUFFLATOR
®
SYSTEM. Use an adjustable or Opened
Ended ¾" wrench to tighten securely. DO NOT USE ANY
TEFLON TAPE OR THREAD SEALING COMPOUNDS.
Reconnect the Yoke to the CO
2
cylinder valve making sure
that the plastic sealing disk is in place.
Open the CO
2
supply tank valve. If CO
2
continues to
prematurely empty or should you hear CO
2
escaping from
any of these connections or the High Pressure Hose itself,
contact US Endoscopy Customer Services.
Disc