CO
2
MPACT ENDOSCOPIC INSUFFLATOR
®
SYSTEM Operator’s Manual
Page 17
Section 9.0
SETTING-UP FOR THE PROCEDURE
9.1 POWER ON AND GAS SUPPLY INDICATOR
1.
If the unit is attached to a wall source central gas supply, switch the Gas Supply
Indicator on the rear panel to “WALL”. Turn on the source supply. Turn on the
Power Switch on the unit. Upon turn on, all lights and indicators will be
illuminated for a brief period.
If a USP Medical Grade gas cylinder is attached to the device, switch the Gas
Supply Indicator on the rear panel to “TANK”. Open the valve on the CO
2
supply tank approximately 1 turn. Turn on the Power Switch on the unit. Upon
turn on, all lights and indicators will be illuminated for a brief period.
2. The GAS SUPPLY INDICATOR LED illuminates red, yellow, or green. The
color coding is a guide to determine whether there is sufficient gas in a supply
tank to finish a procedure, as indicated below:
In Tank Mode
LED Color
Available CO
2
Gas
Green
Tank Pressure is greater than 280 psi
Yellow
Tank Pressure is between 143-280 psi
Red
Tank Pressure is less than 143 psi. Change tank!
NOTE:
If CO
2
supply tank pressure is less than 46 psi (red), flow cannot be initiated.
Flow cannot be initiated in Tank mode if the gas pressure is less than 46 psi.
In Wall Mode
LED Color
Available CO
2
Gas
Green
Central Wall Supply is greater than 30 psi.
Red
Central Wall Supply is less than 30 psi. Flow
cannot be initiated in this state.