Section 8.0
ASSEMBLY PRIOR TO USE
IMPORTANT
If at any time the unit performs erratically or provides otherwise abnormal
operation, remove the unit from service and have it inspected or repaired.
The CO
2
MPACT ENDOSCOPIC INSUFFLATOR
®
SYSTEM should be
inspected upon receipt and before each use. Damaged equipment should be
removed from service and returned to US Endoscopy for repair or replacement.
Before each use, perform the procedures and inspections described in Sections
8.1, 8.2, 8.3, and 8.4.
8.1 PREPARATION
1. Install the CO
2
MPACT ENDOSCOPIC INSUFFLATOR
®
SYSTEM on the
Accessory Cart that can be purchased separately, or on a flat surface, away from
potential sources of spraying or leaking liquids.
2. Position the CO
2
MPACT ENDOSCOPIC INSUFFLATOR
®
SYSTEM so that
the mains power cord connector can easily be reached.
3. The ideal mounting position is even with or above the height of the water bottle.
If the unit is mounted below the water bottle, please use care not to allow water
to siphon from the bottle back into the tubing set.
4. Visually inspect the CO
2
MPACT ENDOSCOPIC INSUFFLATOR
®
SYSTEM
for external signs of damage.
8.2 ELECTRICAL CONNECTIONS
1. Inspect the electrical connections. Do not use if inspection reveals any damage.
2. Connect the Power Cord to the AC Power Connection on the back of the
CO
2
MPACT ENDOSCOPIC INSUFFLATOR
®
SYSTEM.
3. Before connecting the power cord to the hospital grade wall outlet, make sure
that the main power switch is off and that the voltage is correct. The CO
2
MPACT
ENDOSCOPIC INSUFFLATOR
®
SYSTEM has a universal AC line input
device, the nominal AC line voltage is 100 to 240 VAC and the AC line
frequency is 50/60 Hz. The AC line voltage should not drop below 90 VAC or
exceed 264 VAC. There are no switches or other AC line configuration
requirements.