8.3 CONNECTING TO A CO
2
GAS CYLINDER
1. A High Pressure Hose & Yoke (REF 710603) or DIN 477 High Pressure Hose
(REF 390513) is required for connecting to a CO
2
gas cylinder. Contact your
local sales representative or customer service at 1-800-769-8226 to determine
which type of cylinder connection is required in your market.
2. Refer to the Instructions For Use provided with the High Pressure Hose for
further information on assembly and connection to a CO
2
cylinder.
3. Remove cap from the CO
2
Input port on back of the unit. Tighten the other end of
the High-Pressure Hose to the CO
2
Input port on the back of the Unit using a
9/16″ open-ended wrench (see Figure 8.1).
WARNING: Do not allow liquid CO
2
to enter the unit. This can be prevented by assuring that
the CO
2
supply tank is maintained in a vertical upright position at all times.
USE ONLY MEDICAL GRADE CO
2
SIZE “C”, “D”, OR “E” supply tanks.
Tank designations vary by country. Please check with your local distributor for
compatibility.
Figure 8.1
NOTE:
Do not use any teflon tape or
thread sealing compounds on any
connection.