MPACT ENDOSCOPIC INSUFFLATOR
®
SYSTEM Operator’s Manual
Page 13
Section 7.0
REAR PANEL CONTROLS
[7]
GAS INPUT PORT
For connecting a CO
2
supply tank or wall source.
WARNING:
Do not allow liquid CO
2
to enter the unit. This can be prevented by
assuring that the CO
2
supply tank is maintained in a vertical position at all times.
[8]
GAS SOURCE INDICATOR
This switch allows the user to select from a high pressure CO
2
tank or a central supply
source wall connection site.
[9]
AC POWER CONNECTION
Universal AC line input device – nominal AC line voltage 100 to 240 VAC, frequency
50/60 Hz. The AC line voltage should not drop below 90 VAC or exceed 264 VAC.
There are no switches or other AC line configuration requirements.
The line cord (mains lead) supplied with this unit is designed and approved for use in the
USA and Canada only, and should not be used outside these countries. For use outside of
the USA and Canada, your Distributor will supply a line cord that is approved for use in
your country.
AC input is with a standard hospital grade power cord. Connection should be to hospital
grade receptacles only.