9.5 SYSTEM SET-UP
The following operational sequence describes the system set-up for both the
CO
2
MPACT ENDOSCOPIC INSUFFLATOR
®
SYSTEM and Tubing Set:
i)
Make sure the CO
2
MPACT ENDOSCOPIC INSUFFLATOR
®
SYSTEM
is powered up with no Tubing Set in place.
ii)
Make sure the CO
2
MPACT ENDOSCOPIC INSUFFLATOR
®
SYSTEM
is in STOP mode. The green LED will not be illuminated.
iii)
Remove the Tubing Set from its packaging.
iv)
Connect Tubing Set fitting to the endoscopic system.
v)
Connect the Tubing Set fitting to CO
2
MPACT ENDOSCOPIC
INSUFFLATOR
®
SYSTEM.
vi)
Select a flow rate of Low, Medium, or High using the FLOW button.
vii)
Select a shut off time of 30, 60, or 120 Minutes using the TIMER button.
viii)
Start the endoscopic system by turning on the necessary light and video
equipment.
ix)
Verify that the air supply from the endoscopic system is turned off.
x)
Start CO
2
MPACT ENDOSCOPIC INSUFFLATOR
®
SYSTEM by
pressing the START/STOP button.
xi)
Stop the unit at conclusion of procedure.
NOTE:
If an emergency should arise whereby there is a need to terminate the
procedure, the operator should stop the CO
2
gas flow by promptly disconnecting
the Tubing Set at the CO
2
MPACT ENDOSCOPIC INSUFFLATOR
®
SYSTEM
Output port.