![US endoscopy Co2mpact Endoscopic Insufflator Operator'S Manual Download Page 12](http://html.mh-extra.com/html/us-endoscopy/co2mpact-endoscopic-insufflator/co2mpact-endoscopic-insufflator_operators-manual_1063370012.webp)
CO
2
MPACT ENDOSCOPIC INSUFFLATOR
®
SYSTEM Operator’s Manual
Page 12
Section 6.0
FRONT PANEL CONTROLS
NOTE
: The appearance of your CO
2
MPACT ENDOSCOPIC
INSUFFLATOR
®
SYSTEM
may
differ slightly from the units shown in the illustrations and photographs.
[1]
GAS SUPPLY INDICATOR
Tri-color LED to indicate the relative amount of gas available in a
CO
2
supply tank.
[2]
FLOW RATE
Membrane type button that selects the rate of flow of CO
2
that is being insufflated from
the unit. An associated green LED will illuminate to indicate the flow rate. The Low
setting indicates a flow rate of 2.0 Liters/Minute. The Medium setting indicates a flow
rate of 2.9 Liters/Minute. The High setting indicates a flow rate of 3.4 Liters/Minute.
[3]
FLOW TIMER
Membrane type button that selects the amount of time that elapses before the flow of CO
2
automatically stops. An associated green LED will illuminate to indicate the time setting.
The three settings are 30 Minutes, 60 Minutes, and 120 Minutes.
NOTE: The device will default to 60 Minutes when powered on.
[4]
FLOW START/STOP
Membrane type button that starts and stops the flow of CO
2
. An associated green LED
will illuminate to indicate when CO
2
is flowing. The START/STOP button can be
depressed at any time to stop the flow of CO
2
.
NOTE: Flow cannot be initiated in Wall Source mode if the gas pressure is less than 50
psi. Flow cannot be initiated in Tank mode if the gas pressure is less than 46 psi.
[5] GAS OUTPUT CONNECTION TO ENDOSCOPE
Connector for Tubing Sets.
[6]
POWER SWITCH
Turns the power on and off to the unit.