MPACT ENDOSCOPIC INSUFFLATOR
®
SYSTEM Operator’s Manual
Page 29
Condition
Possible Causes
Field Checkout
My CO
2
MPACT
ENDOSCOPIC
INSUFFLATOR
®
SYSTEM does not power-
up at all.
What should I do?
No power at wall outlet.
Blown fuse.
Check CO
2
MPACT ENDOSCOPIC INSUFFLATOR
®
SYSTEM in wall outlet known to be operational.
Remove power cord from rear panel of CO
2
MPACT
ENDOSCOPIC INSUFFLATOR
®
SYSTEM and use a small
screwdriver to open fuse holder.
Pull fuse holder down and examine fuses.
If blown, replace with 2.0-Amp 250-Volt Time-Delay Fuses.