Capacity Setting
21
Chapter
4
4-1.
UTM
Ⅱ
capacity setting procedures (torque)
Perform capacity setting by following the procedures below.
(This cannot be performed when UTM
Ⅱ
is not connected.)
4
UTM
Ⅱ
Capacity Setting
(3) UTM
Ⅱ
capacity setting
*
*
... Be sure to perform the setting. Failure to do so will result in incorrect torque value,
(1) UTM
Ⅱ
capacity
*
(2) Zero calibration
*
rotation speed and driving force value being displayed.
(8) Minimum input
(9) Rotation stop mode
Selecting the UTM
Ⅱ
to perform capacity setting is required
only when there are two UTM
Ⅱ
connected.
Create an unloaded status (where no force etc. is applied to a
sensor), and set the zero point of the sensor.
Select a type of the UTM
Ⅱ
connected.
When there is no type, select "----".
*1: Only when the selected UTM
Ⅱ
type is "----".
Enter the rated output value of the sensor.
Enter the rated capacity value of the sensor.
Select the pulse rate.
Select whether or not to use the absolute value of the torque.
Select minimum input rotation speed.
Select rotation stop mode.
(4) Rated output value
(5) Rated capacity value
(6) Pulse rate *1
(7) Absolute value mode
setting target selection
(V) *1
(Nm) *1
rotation speed
However, (1) applies only when two UTM
Ⅱ
units are connected.