Calibration
34
Calibrat
ion
Chapter
4
4-4.
Display setting of rotation speed/power
Settings related to display of the rotation speed and power are performed.
(1) Pulse rate (rotation speed) setting
Select the pulse input rate.
Select the number of pulses per rotation.
<Setting value>
1 pulse, 4 pulses
(2) Min. scale division (rotation speed) setting
Select the min. scale division (display resolution, scale
division) of the rotation speed.
<Setting value>
1, 2, 5, 10
(3) Unit (power) setting
Select the unit of power.
The decimal place is not automatically changed.
<Setting value>
mW, W, kW, PS, HP
(4) Decimal place (power) setting
Select the decimal place of the power.
The unit is not automatically changed.
<Setting value>
99999, 9999.9, 999.99, 99.999
(5) Min. scale division (power) setting
Select the min. scale division (display resolution, scale
division) of the power.
<Setting value>
1, 2, 5, 10
The power is calculated according to the internal conversion based on the unit
and decimal place setting. Therefore, do not change these settings without due
consideration. Perform re-calibration to change these settings.
Refer to P.37 "4-6.Unit conversion table" for unit conversion table.