Setting Procedures
13
Setting Procedures
Chapter
3
(2) Waveform acquisition mode
Waveform acquisition method is specified.
<Setting value>
Single/repeat/level trigger
Single:
Once the "Start" button is clicked, waveform data acquisition starts,
and waveform data is acquired until the acquisition time is reached.
Repeat:
Once the "Start" button is clicked, waveform data acquisition starts,
and waveform data is acquired until the acquisition time is reached.
Thereafter, waveform data acquisition resumes and is repeated until
the "Stop" button is clicked. Since the file is saved after the waveform data
is acquired and before the next acquisition is started, data is not recorded
during this time.
Level trigger:
When the waveform data specified at the level trigger data in (4) for
trigger, intersects the direction of level values in (5) specified at the slope
in (6), waveform data acquisition starts and the acquisition continues until
the acquisition time is reached.
(3) Pre-trigger
This is enabled when the level trigger is selected in waveform acquisition mode. Waveform data
acquisition time before the trigger point is set in %.
<Setting value>
0 to 99 [%]
(4) Level trigger data
This is enabled when the level trigger is selected in waveform acquisition mode. Waveform data
applicable to the trigger is specified.
<Setting value>
Torque (Torque)/rotation speed (Speed)/power (Power)
(5) Level
This is enabled when the level trigger is selected in waveform acquisition mode. Trigger level is set.
<Setting value>
-99999 to 99999
The setting value unit and decimal places are linked with the
selection in the level trigger data.
(6) Slope
This is enabled when the level trigger is selected in waveform acquisition mode. The intersecting
direction the trigger level is specified.
<Setting value>
Either/positive/negative
(7) Repeat trigger
This is enabled when the level trigger is selected in waveform acquisition mode. When repeat
trigger is checked, the mode turns to state of standby for the trigger condition again after waveform
acquisition. When the check is removed, waveform acquisition is performed once and is complete.