Setting Procedures
14
Setting Procedures
Chapter
3
(8) "Start", "Stop" buttons
"Start" button
The "Start" button is enabled when the display mode is "Normal". The button cannot be clicked
during waveform acquisition or in state of standby for the trigger condition. If waveform
acquisition mode is "Single" or "Repeat", waveform data is acquired from the TM201 and
waveforms are successively displayed when the "Start" button on the bottom right of the screen is
clicked. If waveform acquisition mode is "Level Trigger", it turns to state of standby for the trigger
condition. Once trigger condition is satisfied, waveform display starts.
"Start" button color
If waveform acquisition mode is "Level
trigger", the start button color changes. The
"Start" button remains yellow while
waveform data for the pre-trigger time is
being received after the "Start" button is
clicked.
Waveform data for the pre-trigger time is
received and the "Start" button turns green
once it turns to state of standby for the trigger
condition. However, when the pre-trigger is
0%, the button turns green immediately.
When the trigger conditions are satisfied,
waveform data acquisition starts and the
button turns red.
When data acquisition is complete, the button turns to gray, and in Repeat Trigger setting, it
turns to yellow as in pre-trigger.
After data acquisition, waveform (CSV) files are automatically saved. A maximum of 1000 files
can be saved while old files can be automatically deleted. Refer to P.19 "Automatic deletion of
waveform (CSV) files" of the setting tab.
"Stop" button
This button is enabled when the display mode is "Normal" and during waveform acquisition or
during state of standby for the trigger condition. When the "Stop" button is clicked while waveform
acquisition is in progress, the acquisition stops. When the "Stop" button is clicked during state of
standby for the trigger condition, the wait state is canceled.
During state of standby for the trigger condition or waveform acquisition, operations and settings
other than the "Stop" button cannot be performed. In order to perform graph setting and to switch to
other tabs, the "Stop" button needs to be clicked.
When the "Stop" button is clicked during waveform acquisition, the values after this point in time
are saved as "0" in waveform (CSV) file.
Yellow (pre-trigger in progress)
Green (state of standby for the trigger condition)
Red (data acquisition in progress)