10
2
INSTALLATION & CONNECTION
10
INSTALLATION & CONNECTION
Chapter
2
■
Analog input/output connector connection
Connect a strain gauge sensor or voltage output.
Make connections according to the silk-screen printing on the rear panel of the F350.
●
Pin Assignment
■
How to remove the terminal block
1.
Loosen the screws (two) with a screwdriver.
2.
Remove the terminal block by giving it a
strong pull.
12
SHIELD
11
-MONITOR(ch2)
10
+MONITOR(ch2)
9
-MONITOR(ch1)
8
+MONITOR(ch1)
7
-SIG(ch2)
6
+SIG(ch2)
5
-SIG(ch1)
4
+SIG(ch1)
3
-EXC*
2
+EXC*
1
SHIELD
* EXC is common to ch1 and ch2.
Compatible connector
:
ETB42-12P(OSADA CO.,Ltd.) or equivalent
(Connector optional type
:
CN81)
CAUTION
Before connecting a strain gauge sensor, make sure to check the setting of excitation
voltage.
For the setting of excitation voltage, see "4-2
Excitation voltage setting" on page
Loosen.
(Turn
counterclockwise
.)
When installing the terminal block to the F350
body, check its orientation. (See the illustration at
the right.)
Attention
Insert side
Right side