Operating Manual for UTD2000/3000
83
Figure 3-9 Trigger of video row
It has to check if input signal is within rule scope or
not. Fail refers to out of scope and Pass refers to
within the scope. Pass/Fail signal can be shown by
output port
.
1. To press[UTILITY] button then press[F2]Enter into
Pass/Fail menu.
2. Setup of information source: Enter into Pass/Fail
menu then press[F2] to select information source.
3. Template setup: To press[F4] Enter into template
setup menu (Template). To press[F2] and[F3] shown
in Template menu to set horizontal and vertical
tolerance scope by multi-function knob, (namely, 1-
200Pixel toward horizontal direction and 1-100Pixel
toward vertical direction.) Please press F4 finally
Return to Pass/Fail menu.
4. Setup of output conditions: To press[F3] to set
output conditions shown in Pass/Fail menu.
5. To begin detection: To press[F1] to open
detection status shown in Pass/Fail menu.
The eighth example: Pass/Fail detection
of the rear panel
Rear panel of Pass / Fail output
Pass FailInterface Schematic