79
Ƹ
V
namely
pressure difference can be shown
in cursor menu automatically
Please refer to following figure for details.
Note: Please set cursor type to time if onlying
adopting the secondstep as above for time
measurement by cursor.
Figure 3-6 Voltage difference of signal by cursor
measurement
To review phase location difference of signals of 2
channels.
Example: Phase location variation caused by 1
circuit for tested signals. Please connect digital
storage oscillograph to circuit to monitor
input/output signal of circuits. Please operate
according to following steps to review input/output
terminal of circuits by way of X-Y coordinate
diagram.
1. To set probe menu attenuation coefficient to 10*
and set switch on the probe to 10*.
2. To connect CH1 probe to network input then
connect Ch2 probe to network output.
3. To press CH1 and CH2 menu keys and open 2
channels if they have not been displayed.
4. To press AUTO button.
5. To adjust vertical scale knob for approximate
equivalent display amplitude of 2 loops of signals.
The sixth example: Application of X-Y function
Operating Manual for UTD2000/3000