METHOD
Stage 1 - Preparation of the Test System
Starting up the Reflotron®
•
Ensure that the Reflotron® is set up away from direct sunlight and sources of
heat (radiators, heaters or instruments that radiate heat). Room temperature
should be b15
0
C and +34
0
C.
•
Plug the power cable into the back of the Reflotron® and then into a normal
240v wall socket.
•
Switch power on at the wall.
•
Switch Reflotron® to “ON” by pressing the
power switch on the front of the instrument.
•
The Reflotron® will then warm up and
conduct some self-checks before
displaying the message “READY”.
NB. Never obstruct the vents or fan outlets on the instrument.
Daily Maintenance
a) Clean and Check Strip
The check strips are special strips
which are used to check the optical
system.