What to do if the Reflotron® QC results are unacceptable
QC FAIL
(Readings out of range)
YES
OK to resume
testing
Call Customer
Support
NO
Check expiry dates
of QC material &
test strips
Are they in date?
Perform cleaning
procedure (Page26)
Repeat the QC
Readings in range?
YES
NO
Discard expired
components &
replace
Repeat the QC
Readings in range?
YES
NO
OK to resume
testing
Reconstitute a
new bottle of
QC material
Repeat the QC
Readings in
range?
YES
NO