Running the controls
Preparing the Reflotron® test strip
•
Remove the appropriate strip from its container and replace lid
•
Place strip on a clean dry surface
Adding the control
•
Using the dedicated 32µl pipette and a clean dry tip, dispense 32µl of
reconstituted control onto the appropriate test strip
•
Check there are no bubbles in the tip
•
Hold the pipette and tip vertically over the sample
pad on the test strip
•
Follow the steps as for patient sample –Page 32 to
33 Steps 4 – 5
Reviewing the Results
•
Check the result against the values provided in the appropriate pack insert
•
If the control is within the acceptable range, record the result and continue
testing
•
If the control is outside the acceptable range, record the result and repeat the
test