The Test Strips
•
No preparation required
•
Storage conditions – most at RT (see Appendix 2 for individual tests)
•
The test strip containers have desiccants contained in the lids to keep the
strips dry. The lid must be replaced after every opening.
•
Once the strip has been removed it must be used within the hour
•
Once the blood sample has been added , the test strip must be tested
immediately
•
Shelf life – 6 – 12 months (see expiry date on strip container)
•
Must be disposed of as clinical waste
Calibration
•
No calibration required – all calibration information is stored on the magnetic
strip on the reverse of the test strip
Sample Pad
Reagent
Layers
Magnetic strip
located on the
underside
of
the test strip