INTERNAL QUALITY CONTROL (IQC)
(All information, instructions and pictures referring to the
Reflotron®® A1C Quality Control are taken from the QC
pack insert)
Running a Quality Control (QC)
The QC should be assayed in the same manner as patient samples
As for clinical specimens, all QC samples should be handled as if capable of
transmitting infection. Appropriate procedures should be used for their disposal.
The Reflotron® Quality Control (QC) Kit
The Reflotron® Quality Control should be used with the Reflotron® test strips to
ensure that the strips are working correctly and giving accurate and reliable results.
Precinorm Hb QC kit should be used with
the Hb test strips
•
4 x 2ml bottles for reconstitution with
2mls deionised/distilled water
•
4 x level 1
Precinorm HDL QC kit should be used with
the HDL test strips
•
4 x 2ml bottles for reconstitution with
2mls deionised/distilled water
•
2 x level 1, 2 x level 2
Precinorm U QC kit should be used for all other test strips
•
4 x 2ml bottles for reconstitution with 2mls deionised/distilled water
•
4 x level 1