Running a Clean & Check Strip
•
Open the container and remove one Check Strip, replace lid
•
Open the lid of the measuring chamber and insert the
check strip. Close the measuring chamber lid
•
The Reflotron® will now read the magnetic strip on the
underside of the Check Strip and will automatically begin
the test
•
The Check takes approximately 1 minute
,
•
At the end of the check time, three values will be displayed. These values
must be as close as possible to the middle row of numbers on the label of the
check strip vial, and definitely between the ‘maximum’ and ‘minimum’ values.
•
Record the results